Development and
Validation of RP-HPLC Method for the Estimation of Bilastine from bulk and
Formulation
Pardeshi P. P.*, Gaware V. M.,
Dhamak K. B.
Department of Quality
Assurance, Pres’s College of Pharmacy (For Women), Chincholi, Tal: Sinnar,
Dist: Nashik (MS) India.
*Corresponding Author E-mail: payalp0220@gmail.com
ABSTRACT:
A Simple, linear, precise,
accurate, robust and selective RP-HPLC method has been developed for the
estimation of Bilastine from bulk and tablet formulation. The methanol:
orthophosporic acid buffer in proportion of (70:30) was used as mobile phase.
The HPLC system consisting of LC20AD Prominence Liquid Chromatography SPD 20-A
Shimadzu Japan. The UV-VIS detector and C18 column with dimension on 250×4.6 mm
was used at wavelength 280 nm. Finally Bilastine was quantified from bulk and
its tablet formulation. The proposed method was found to be linear, precise and
accurate as per the ICH guidelines.
KEYWORDS: Bilastine, RP-
HPLC, Validation, Method development.
INTRODUCTION:
Bilastine is second generation
anti-histamine used in treatment of allergic
rhinoconjunctivitis and urticaria. Chemically it is (3S, 4R)-3-Methyl-4-phenyl-1-[1s,
4s)-4-cyano-4-(4-fluorophenyl) cyclohexyl] piperidine-4-carboxylic acid. [1,2]
Fig. Structure of Bilastine
Pharmacological
studies have shown that the drug is highly selective for the H1 receptor in
both in vivo and in vitro studies, and with no apparent affinity
for other receptors. The absorption of Bilastine is fast, linear and dose
proportional; it appears to be safe and well tolerated at all doses levels in
healthy population.[1]
Several methods had
been reported for the estimation of Bilastine but not a single method is
available with the utilization of mobile phase which less harmful to HPLC
column. Hence in the view of that present method had been developed by using
suitable mobile phase.
EXPERIMENTAL:
Bilastine chemical
reference standard was supplied by SYMED LABS LIMITED (UNIT-I), Telangana. The
remaining chemicals were purchased from modern science Nashik and all are HPLC
grade.
HPLC
(High-Performance Liquid Chromatography) [5-10]:
HPLC method was
carried out for Bilastine bulk and formulation. The LC20AD Prominence Liquid
Chromatography SPD20-A Shimadzu, Japan with UV-Vis detector and C18 column with
dimension on 250×4.6 mm was used for the method development with flow rate 0.8
ml/min at wavelength 280 nm. The methanol:
Orthophosphoric
acid buffer in proportion of (70v:30v) was used as a mobile phase, for
development of chromatogram. The method was validated for Bilastine and various
parameters according to ICH guidelines were studied.
Stationary Phase:
C18 column with
dimension on 250×4.6 mm are used.
Preparation of
Mobile Phase:
The selection of
mobile phase was according to polarity and non-polarity of solvents. The
methanol: Orthophosphoric acid buffer was selected as mobile phase in ratio of
70:30 and was filtered on membrane filter (0.45μ) to remove degassing and
were stirred for 10-15 min. [3,4]
Preparation of
Stock Solution (Standard):
The stock solution
of 100μg/ml was prepared by dissolving 10 mg Bilastine in 100 ml mobile
phase. The dilution was prepared in various concentrations using stock solution
and was diluted in mobile phase.
Preparation of
Sample Solution (Formulation):
The sample solution
of Bilastine formulation was prepared as 100μg/ml stock solution for
quantification of Bilastine in tablet formulation. The dilution was prepared in
various concentrations using sample stock and was dissolved in mobile phase for
quantification of Bilastine formulation.
System Suitability
Parameters:
The area of
respective concentrations, theoretical plates, number of theoretical plates per
cm, Tailing factor and the peak symmetry was recorded. [5,6]
Table no. 01:
Optimized chromatographic conditions for RP-HPLC
|
Chromatographic conditions |
SHIMADZU HPLC System |
|
Mobile Phase |
Methanol: Orthophosphoric acid buffer(70v:30v) |
|
Column |
ARP-C18 (250mm×4.6 mm), 5μ column |
|
Flow rate |
1.0 ml/min |
|
Wavelength detection |
280 nm |
|
Injection volume |
10µl |
|
Temperature |
Room temperature |
|
Retention time |
3.280 |
|
Run time |
40 minutes |
Linearity:
Dilution of
standard Bilastine in the range of 5-30μg/ml were prepared by taking
suitable aliquots of working standard solution in different 10ml volumetric
flasks and diluting up to the mark with mobile phase. 10μl was injected
from it each time on column at flow rate of 1.0 ml/min. The standard from elute
was monitored at 280 nm and corresponding chromatogram were obtained from these
chromatograms peak area were calculated. A plot of peak area over concentration
was constructed. Regression of the plot was computed by least square regression
method. [5,6,7]
Fig. no. 02:
Chromatogram for optimized method
Precision:
Precision of
analytical method was studied by multiple injections of homogeneous samples. 6
replicate of 10 ppm solution were prepared and injected for precision at the
same flow rate of 1.0 ml/min. The inter-day and intermediate precisions were
used to study the variability of the method S.D. and %R.S.D. was calculated for
both.
Robustness:
Robustness was
studied by changing parameters like change in flow rate. The S.D. and %R.S.D.
between the change parameter were calculated.
Ruggedness:
Ruggedness study
was carried out by using different analysts. The S.D. and %R.S.D. were
calculated.
Recovery:
Recovery of the
method was studied using the method of standard addition. Standard Bilastine
solutions were added to the unknown bulk and tablet formulation of Bilastine.
The percent recovery was determined at three different levels (50%, 100%, and
150%). Impurity content was determined and the percent recovery was calculated.
]Table No. 02:
Method Validation Parameters of HPLC
|
Sr. No. |
Parameter |
Observation |
|
1 |
Linearity range |
5-30μg/ml |
|
2 |
Slope |
14.97 |
|
3 |
Intercept |
48.96 |
|
4 |
Correlation coefficient |
0.997 |
|
5 |
LOD |
0.2177μg/ml |
|
6 |
LOQ |
0.6597μg/ml |
Table no. 03:
Summary of Repeatability studies
|
Sr. No. |
Parameter |
SD |
%RSD |
|
1 |
Precision |
0.987 |
0.824 |
|
2 |
Intraday precision |
0.901 |
0.755 |
|
3 |
Interday Precision |
1.279 |
1.046 |
|
4 |
Robustness |
1.970 |
0.497 |
|
5 |
Ruggedness |
0.831 |
0.690 |
The summary of
repeatability studies was found to be ≤2
Table no. 04: Result
of recovery study by HPLC
|
Sr. No. |
Drug/ Formulation |
Percentage recovery |
Mean |
SD |
%RSD |
||
|
|
|
50% |
100% |
150% |
|
|
|
|
1 |
Bulk |
97 |
98.62 |
99.2 |
98.27 |
1.140 |
1.160 |
|
2 |
Tablet |
97.83 |
98.87 |
99.5 |
98.73 |
0.843 |
0.854 |
Table No. 05: System
suitability Parameters
|
Sr. No. |
Property |
Values |
Official limits |
|
1 |
Retention time (tR) |
3.28 |
- |
|
2 |
Theoretical plate (N) |
4675 |
N≥2000 |
|
3 |
Resolution (R) |
4.652 |
R≥2 |
|
4 |
Tailing factor (T) |
0.98 |
T≤2 |
LOD and LOQ:
Limit of detection
and limit of Quantitation of the method was calculated by formula given below,
LOD = 3.3 ×
S.D/Slope
LOQ =10 ×
S.D./Slope
CONCLUSION:
The proposed method
was simple, rapid and selective. The percentage relative standard deviation was
found to be below 2.0% which indicate method was highly precise and accurate.
The retention time of Bilastine by HPLC method was found to be 3.28 minutes the
amount of Bilastine present in bulk was found to be 98.85% and in tablet was
found to be 99.25% the above method was used for routine analysis.
REFERENCES:
1. T. Zuberbier, C.
Bindslev-Jensen, W. Canonica et al., “EAACI/ GA2LEN/EDF guideline: management
of urticaria,” Allergy, vol. 61, no. 3, pp. 321–331, 2006.
2. ICH Q2B: Guidance
for Industry: Validation of Analytical Procedures: Methodology U.S. Department
of Health and Human Services Food and Drug Administration, (CDER), (CBER),
1996, 1-10.
3. F. Settle, Handbook
of Instrumental Techniques for Analytical Chemistry, Pearson Education, First
Indian Reprint 2004, 147-160.
4. L.R. Snyder, J.J.
Kirkland, J.L. Glajch, Practical HPLC Method Development, Wiley Interscience,
2, 31-42.
5. David G. Watson,
Pharmaceutical Analysis, A Text book for pharmacy students & pharmaceutical
chemists; Churchill Livingstone Publications, 1999; 197-199.
6. Michael Swartz:
Analytical method development and validation, 2009, Page no: 25-27, 60-67.
7. Chung Chow Chan,
Herman Lam.: Analytical method 444validation and instrument performance
verification, Page no: 37, 41, 43, 47.
Received on 09.03.2020 Modified on 19.03.2020
Accepted on 08.04.2020
©Asian Pharma Press All Right Reserved
Asian J. Pharm. Ana. 2020; 10(2):109-111.
DOI: 10.5958/2231-5675.2020.00019.8