Development and Validation of RP-HPLC Method for the Estimation of Bilastine from bulk and Formulation

 

Pardeshi P. P.*, Gaware V. M., Dhamak K. B.

Department of Quality Assurance, Pres’s College of Pharmacy (For Women), Chincholi, Tal: Sinnar,

Dist: Nashik (MS) India.

*Corresponding Author E-mail: payalp0220@gmail.com

 

ABSTRACT:

A Simple, linear, precise, accurate, robust and selective RP-HPLC method has been developed for the estimation of Bilastine from bulk and tablet formulation. The methanol: orthophosporic acid buffer in proportion of (70:30) was used as mobile phase. The HPLC system consisting of LC20AD Prominence Liquid Chromatography SPD 20-A Shimadzu Japan. The UV-VIS detector and C18 column with dimension on 250×4.6 mm was used at wavelength 280 nm. Finally Bilastine was quantified from bulk and its tablet formulation. The proposed method was found to be linear, precise and accurate as per the ICH guidelines.

 

KEYWORDS: Bilastine, RP- HPLC, Validation, Method development.

 

 

 

INTRODUCTION:

Bilastine is second generation anti-histamine used in treatment of allergic rhinoconjunctivitis and urticaria. Chemically it is (3S, 4R)-3-Methyl-4-phenyl-1-[1s, 4s)-4-cyano-4-(4-fluorophenyl) cyclohexyl] piperidine-4-carboxylic acid. [1,2]

 

 

Fig. Structure of Bilastine

 

Pharmacological studies have shown that the drug is highly selective for the H1 receptor in both in vivo and in vitro studies, and with no apparent affinity for other receptors. The absorption of Bilastine is fast, linear and dose proportional; it appears to be safe and well tolerated at all doses levels in healthy population.[1]

 

Several methods had been reported for the estimation of Bilastine but not a single method is available with the utilization of mobile phase which less harmful to HPLC column. Hence in the view of that present method had been developed by using suitable mobile phase.

 

EXPERIMENTAL:

Bilastine chemical reference standard was supplied by SYMED LABS LIMITED (UNIT-I), Telangana. The remaining chemicals were purchased from modern science Nashik and all are HPLC grade.

 

HPLC (High-Performance Liquid Chromatography) [5-10]:

HPLC method was carried out for Bilastine bulk and formulation. The LC20AD Prominence Liquid Chromatography SPD20-A Shimadzu, Japan with UV-Vis detector and C18 column with dimension on 250×4.6 mm was used for the method development with flow rate 0.8 ml/min at wavelength 280 nm. The methanol:

 

Orthophosphoric acid buffer in proportion of (70v:30v) was used as a mobile phase, for development of chromatogram. The method was validated for Bilastine and various parameters according to ICH guidelines were studied.

 

Stationary Phase:

C18 column with dimension on 250×4.6 mm are used.

 

Preparation of Mobile Phase:

The selection of mobile phase was according to polarity and non-polarity of solvents. The methanol: Orthophosphoric acid buffer was selected as mobile phase in ratio of 70:30 and was filtered on membrane filter (0.45μ) to remove degassing and were stirred for 10-15 min. [3,4]

 

Preparation of Stock Solution (Standard):

The stock solution of 100μg/ml was prepared by dissolving 10 mg Bilastine in 100 ml mobile phase. The dilution was prepared in various concentrations using stock solution and was diluted in mobile phase.

 

Preparation of Sample Solution (Formulation):

The sample solution of Bilastine formulation was prepared as 100μg/ml stock solution for quantification of Bilastine in tablet formulation. The dilution was prepared in various concentrations using sample stock and was dissolved in mobile phase for quantification of Bilastine formulation.

 

System Suitability Parameters:

The area of respective concentrations, theoretical plates, number of theoretical plates per cm, Tailing factor and the peak symmetry was recorded. [5,6]

 

Table no. 01: Optimized chromatographic conditions for RP-HPLC

Chromatographic conditions

SHIMADZU HPLC System

Mobile Phase

Methanol: Orthophosphoric acid buffer(70v:30v)

Column

ARP-C18 (250mm×4.6 mm), 5μ column

Flow rate

1.0 ml/min

Wavelength detection

280 nm

Injection volume

10µl

Temperature

Room temperature

Retention time

3.280

Run time

40 minutes

Linearity:

Dilution of standard Bilastine in the range of 5-30μg/ml were prepared by taking suitable aliquots of working standard solution in different 10ml volumetric flasks and diluting up to the mark with mobile phase. 10μl was injected from it each time on column at flow rate of 1.0 ml/min. The standard from elute was monitored at 280 nm and corresponding chromatogram were obtained from these chromatograms peak area were calculated. A plot of peak area over concentration was constructed. Regression of the plot was computed by least square regression method. [5,6,7]

 

Fig. no. 02: Chromatogram for optimized method

 

Precision:

Precision of analytical method was studied by multiple injections of homogeneous samples. 6 replicate of 10 ppm solution were prepared and injected for precision at the same flow rate of 1.0 ml/min. The inter-day and intermediate precisions were used to study the variability of the method S.D. and %R.S.D. was calculated for both.

 

Robustness:

Robustness was studied by changing parameters like change in flow rate. The S.D. and %R.S.D. between the change parameter were calculated.

 

Ruggedness:

Ruggedness study was carried out by using different analysts. The S.D. and %R.S.D. were calculated.

 

Recovery:

Recovery of the method was studied using the method of standard addition. Standard Bilastine solutions were added to the unknown bulk and tablet formulation of Bilastine. The percent recovery was determined at three different levels (50%, 100%, and 150%). Impurity content was determined and the percent recovery was calculated.

 

 

 

]Table No. 02: Method Validation Parameters of HPLC

Sr. No.

Parameter

Observation

1

Linearity range

5-30μg/ml

2

Slope

14.97

3

Intercept

48.96

4

Correlation coefficient

0.997

5

LOD

0.2177μg/ml

6

LOQ

0.6597μg/ml

Table no. 03: Summary of Repeatability studies

Sr. No.

Parameter

SD

%RSD

1

Precision

0.987

0.824

2

Intraday precision

0.901

0.755

3

Interday Precision

1.279

1.046

4

Robustness

1.970

0.497

5

Ruggedness

0.831

0.690

The summary of repeatability studies was found to be ≤2

 

 

 

Table no. 04: Result of recovery study by HPLC

Sr. No.

Drug/ Formulation

Percentage recovery

Mean

SD

%RSD

 

 

50%

100%

150%

 

 

 

1

Bulk

97

98.62

99.2

98.27

1.140

1.160

2

Tablet

97.83

98.87

99.5

98.73

0.843

0.854

 

 

Table No. 05: System suitability Parameters

Sr. No.

Property

Values

Official limits

1

Retention time (tR)

3.28

-

2

Theoretical plate (N)

4675

N≥2000

3

Resolution (R)

4.652

R≥2

4

Tailing factor (T)

0.98

T≤2

 

LOD and LOQ:

Limit of detection and limit of Quantitation of the method was calculated by formula given below,

 

LOD = 3.3 × S.D/Slope

LOQ =10 × S.D./Slope

 

CONCLUSION:

The proposed method was simple, rapid and selective. The percentage relative standard deviation was found to be below 2.0% which indicate method was highly precise and accurate. The retention time of Bilastine by HPLC method was found to be 3.28 minutes the amount of Bilastine present in bulk was found to be 98.85% and in tablet was found to be 99.25% the above method was used for routine analysis.

 

REFERENCES:

1.      T. Zuberbier, C. Bindslev-Jensen, W. Canonica et al., “EAACI/ GA2LEN/EDF guideline: management of urticaria,” Allergy, vol. 61, no. 3, pp. 321–331, 2006.

2.      ICH Q2B: Guidance for Industry: Validation of Analytical Procedures: Methodology U.S. Department of Health and Human Services Food and Drug Administration, (CDER), (CBER), 1996, 1-10.

3.      F. Settle, Handbook of Instrumental Techniques for Analytical Chemistry, Pearson Education, First Indian Reprint 2004, 147-160.

4.      L.R. Snyder, J.J. Kirkland, J.L. Glajch, Practical HPLC Method Development, Wiley Interscience, 2, 31-42.

5.      David G. Watson, Pharmaceutical Analysis, A Text book for pharmacy students & pharmaceutical chemists; Churchill Livingstone Publications, 1999; 197-199.

6.      Michael Swartz: Analytical method development and validation, 2009, Page no: 25-27, 60-67.

7.      Chung Chow Chan, Herman Lam.: Analytical method 444validation and instrument performance verification, Page no: 37, 41, 43, 47.

 

 

 

Received on 09.03.2020       Modified on 19.03.2020

Accepted on 08.04.2020      ©Asian Pharma Press All Right Reserved

Asian J. Pharm. Ana. 2020; 10(2):109-111.

DOI: 10.5958/2231-5675.2020.00019.8